Infant Incubator


eneral

parameters

Model BJPX-3101
Power Supply AC220-230V/50Hz,1300VA
Maximum heater output power 350W/240V
Auxiliary power output AC220-230V/50Hz,3A
Heater instructions 0~100%
Temperature control mode Box temperature mode; infant mode
Box temperature control temperature range 25~37℃;37~39℃(Leapfrogging Mode)
Skin temperature control temperature range 34~37℃37~38℃(Leapfrogging Mode)
Sensor temperature display range 5~65℃
Warm-up time(ambient temperature is generally + 22 ℃)*  

No more than 40 minutes

 

Temperature variability* No more than 0.5℃
Temperature uniformity * (the mattress is in a horizontal position) No more than 0.8℃
Temperature uniformity * (the mattress is in an inclined position) No more than 1.0℃
The difference between average incubator temperature and control temperature No more than 1.0℃
Skin temperature sensor accuracy ±0.2℃
Humidity

control

The running time of the water tank is filled with distilled water after Humidity below 70%RH, at least 24 hours
Tank capacity 1200ml
Humidity display range 0%RH~99%RH
Humidity control range 0%~90%(Adjust 1% each time)
Humidity control precision ±5%RH
Humidity display accuracy

(ambient temperature 25℃, ambient humidity 45%RH + 5% RH)

±5%RH
Note: in the context of a fairly high ambient humidity, it may not be possible to achieve a

relatively low humidity control

Oxygen

concentration

control

Oxygen concentration display range 0%~99%
Oxygen concentration display resolution 1%
Oxygen concentration display accuracy ±2% O?(The oxygen concentration set value of less than25%)

±3% O?(The oxygen concentration set value of more than 25%)

Oxygen concentration control accuracy ±4%O?Concentration within the volume
Oxygen concentration setting range 20%~60%(Adjust 1% each time)
Oxygen sensor service life Use up to 10,000 hours at 100% oxygen concentration
Ambient

temperature

The scope of work +20~+30℃
Transportation and storage -20~+55℃
(suggested that the user should not be used in an environment outside the specified range)
Relative humidity Relative humidity operating range 30%~75%RH
Transportation and storage No more than 93%RH
Note 🙁 not recommended to be used outside the specified range of environments)
Atmospheric pressure Transport and storage atmospheric pressure range 500hPa~1060hPa
Working atmospheric pressure range 800hPa~1060hPa
Applicable Environment Altitude ≤2000m
Overvoltage category
Pollution Level  

2

Air flow rate Ambient air flow rate Less than 0.3m/s
Other indicators Ambient noise in constant temperature enclosure The stable temperature is not more than 45dB (A)

In other states is not more than 50dB (A)

Ambient noise is not more than 35dB (A)

Ambient carbon dioxide concentration in constant temperature enclosure Carbon dioxide gas mixture is injected at the

speed of 750ml/minute right above the mattress about 10cm, the density should be under 0.5%.

Air velocity above the crib mattress No more than0.35m/s
General alarm  

Power off alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm,

sensor box placement error alarm, water tank placement error alarm, water shortage alarm

 

System alarm ROM alarm, internal alarm system,SRAM alarm, EEPROM alarm, real-time alarm,

communication alarm, ADC alarm, keyboard alarm, battery alarm, LCD display alarm,

temperature heating system alarm, humidity heating system alarm

Note: please refer to the contents of the relevant terms and markers.

eneral

parameters

Model BJPX-3101
Power Supply AC220-230V/50Hz,1300VA
Maximum heater output power 350W/240V
Auxiliary power output AC220-230V/50Hz,3A
Heater instructions 0~100%
Temperature control mode Box temperature mode; infant mode
Box temperature control temperature range 25~37℃;37~39℃(Leapfrogging Mode)
Skin temperature control temperature range 34~37℃37~38℃(Leapfrogging Mode)
Sensor temperature display range 5~65℃
Warm-up time(ambient temperature is generally + 22 ℃)*  

No more than 40 minutes

 

Temperature variability* No more than 0.5℃
Temperature uniformity * (the mattress is in a horizontal position) No more than 0.8℃
Temperature uniformity * (the mattress is in an inclined position) No more than 1.0℃
The difference between average incubator temperature and control temperature No more than 1.0℃
Skin temperature sensor accuracy ±0.2℃
Humidity

control

The running time of the water tank is filled with distilled water after Humidity below 70%RH, at least 24 hours
Tank capacity 1200ml
Humidity display range 0%RH~99%RH
Humidity control range 0%~90%(Adjust 1% each time)
Humidity control precision ±5%RH
Humidity display accuracy

(ambient temperature 25℃, ambient humidity 45%RH + 5% RH)

±5%RH
Note: in the context of a fairly high ambient humidity, it may not be possible to achieve a

relatively low humidity control

Oxygen

concentration

control

Oxygen concentration display range 0%~99%
Oxygen concentration display resolution 1%
Oxygen concentration display accuracy ±2% O?(The oxygen concentration set value of less than25%)

±3% O?(The oxygen concentration set value of more than 25%)

Oxygen concentration control accuracy ±4%O?Concentration within the volume
Oxygen concentration setting range 20%~60%(Adjust 1% each time)
Oxygen sensor service life Use up to 10,000 hours at 100% oxygen concentration
Ambient

temperature

The scope of work +20~+30℃
Transportation and storage -20~+55℃
(suggested that the user should not be used in an environment outside the specified range)
Relative humidity Relative humidity operating range 30%~75%RH
Transportation and storage No more than 93%RH
Note 🙁 not recommended to be used outside the specified range of environments)
Atmospheric pressure Transport and storage atmospheric pressure range 500hPa~1060hPa
Working atmospheric pressure range 800hPa~1060hPa
Applicable Environment Altitude ≤2000m
Overvoltage category
Pollution Level  

2

Air flow rate Ambient air flow rate Less than 0.3m/s
Other indicators Ambient noise in constant temperature enclosure The stable temperature is not more than 45dB (A)

In other states is not more than 50dB (A)

Ambient noise is not more than 35dB (A)

Ambient carbon dioxide concentration in constant temperature enclosure Carbon dioxide gas mixture is injected at the

speed of 750ml/minute right above the mattress about 10cm, the density should be under 0.5%.

Air velocity above the crib mattress No more than0.35m/s
General alarm  

Power off alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm,

sensor box placement error alarm, water tank placement error alarm, water shortage alarm

 

System alarm ROM alarm, internal alarm system,SRAM alarm, EEPROM alarm, real-time alarm,

communication alarm, ADC alarm, keyboard alarm, battery alarm, LCD display alarm,

temperature heating system alarm, humidity heating system alarm

Note: please refer to the contents of the relevant terms and markers.