eneral
parameters |
Model |
BJPX-3101 |
Power Supply |
AC220-230V/50Hz,1300VA |
Maximum heater output power |
350W/240V |
Auxiliary power output |
AC220-230V/50Hz,3A |
Heater instructions |
0~100% |
Temperature control mode |
Box temperature mode; infant mode |
Box temperature control temperature range |
25~37℃;37~39℃(Leapfrogging Mode) |
Skin temperature control temperature range |
34~37℃37~38℃(Leapfrogging Mode) |
Sensor temperature display range |
5~65℃ |
Warm-up time(ambient temperature is generally + 22 ℃)* |
No more than 40 minutes
|
Temperature variability* |
No more than 0.5℃ |
Temperature uniformity * (the mattress is in a horizontal position) |
No more than 0.8℃ |
Temperature uniformity * (the mattress is in an inclined position) |
No more than 1.0℃ |
The difference between average incubator temperature and control temperature |
No more than 1.0℃ |
Skin temperature sensor accuracy |
±0.2℃ |
Humidity
control |
The running time of the water tank is filled with distilled water after |
Humidity below 70%RH, at least 24 hours |
Tank capacity |
1200ml |
Humidity display range |
0%RH~99%RH |
Humidity control range |
0%~90%(Adjust 1% each time) |
Humidity control precision |
±5%RH |
Humidity display accuracy
(ambient temperature 25℃, ambient humidity 45%RH + 5% RH) |
±5%RH |
Note: in the context of a fairly high ambient humidity, it may not be possible to achieve a
relatively low humidity control |
Oxygen
concentration
control |
Oxygen concentration display range |
0%~99% |
Oxygen concentration display resolution |
1% |
Oxygen concentration display accuracy |
±2% O?(The oxygen concentration set value of less than25%)
±3% O?(The oxygen concentration set value of more than 25%) |
Oxygen concentration control accuracy |
±4%O?Concentration within the volume |
Oxygen concentration setting range |
20%~60%(Adjust 1% each time) |
Oxygen sensor service life |
Use up to 10,000 hours at 100% oxygen concentration |
Ambient
temperature |
The scope of work |
+20~+30℃ |
Transportation and storage |
-20~+55℃ |
(suggested that the user should not be used in an environment outside the specified range) |
Relative humidity |
Relative humidity operating range |
30%~75%RH |
Transportation and storage |
No more than 93%RH |
Note 🙁 not recommended to be used outside the specified range of environments) |
Atmospheric pressure |
Transport and storage atmospheric pressure range |
500hPa~1060hPa |
Working atmospheric pressure range |
800hPa~1060hPa |
Applicable Environment Altitude |
≤2000m |
Overvoltage category |
Ⅱ |
Pollution Level |
2 |
Air flow rate |
Ambient air flow rate |
Less than 0.3m/s |
Other indicators |
Ambient noise in constant temperature enclosure |
The stable temperature is not more than 45dB (A)
In other states is not more than 50dB (A)
Ambient noise is not more than 35dB (A) |
Ambient carbon dioxide concentration in constant temperature enclosure |
Carbon dioxide gas mixture is injected at the
speed of 750ml/minute right above the mattress about 10cm, the density should be under 0.5%. |
Air velocity above the crib mattress |
No more than0.35m/s |
General alarm |
Power off alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm,
sensor box placement error alarm, water tank placement error alarm, water shortage alarm
|
System alarm |
ROM alarm, internal alarm system,SRAM alarm, EEPROM alarm, real-time alarm,
communication alarm, ADC alarm, keyboard alarm, battery alarm, LCD display alarm,
temperature heating system alarm, humidity heating system alarm |
Note: please refer to the contents of the relevant terms and markers. |